With barcode reader, electronic zero, vial rotation, average function
The HI802 automatically identifies barcoded sample vials, detecting method type and range to minimize errors and streamline measurements.
Vial rotation during measurement enables method identification and multiple absorbance readings, which are converted to concentration units and displayed on the LCD. This rotational averaging improves method accuracy.
Single-zero measurement with DI water blank correction ensures stable readings and reflects sample features, not instrument changes, for ease of use.
All commercially available spectrophotometers take a single reading, except for one competitor, which takes an average reading based on 10 absorbance readings. Hanna Instruments has raised the average to 256 readings!
By submitting this form, I agree to the Terms and Conditions and Privacy Policy available on the Hanna Instruments Australia website.
The HI802 comes with over 100 pre-programmed methods for chemical analysis. Methods can be updated via computer or flash drive, and up to 150 can be saved. It also provides reagent codes for easy ordering.
Program up to 100 custom methods with up to 10 calibration points, 5 simultaneous wavelengths, and 5 reaction timers. This flexibility allows for extensive customization beyond factory presets, accommodating specific needs and modifications.
29 methods with pre-dosed vials with barcode
Automatic method recognition
Up to 150 factory methods
104 methods pre-programmed
Updateable via USB key
Up to 100 user methods
The tool can be customised to suit your needs
Possibility of creating a new method with the AVERAGE reading function
Quick access to the most used methods
| Parameter | Range | Resolution | Accuracy | Wavelength | Method | Reagent Code |
| Alkalinity | 0 to 500 mg/L (as CaCO) | 1 mg/L | ±5 mg/L ±5% of reading at 25 °C | 610 nm | Bromocresol Green | HI775-26 Reagents for 25 tests |
| Alkalinity, Marine | 0 to 300 mg/L (as CaCO) | 1 mg/L | ±5 mg/L ±5% of reading at 25 °C | 610 nm | Bromocresol Green | HI755-26 Reagents for 25 tests |
| Aluminum | 0.00 to 1.00 mg/L (as Al³+) | 0.01 mg/L | ±0.04 mg/L ±4% of reading at 25 °C | 530 nm | Adaptation of the Aluminon Method | HI93712-01 Reagents for 100 tests; HI93712-03 Reagents for 300 tests |
| Ammonia Low Range | 0.00 to 3.00 mg/L (as NH-N) | 0.01 mg/L | ±0.04 mg/L ±4% of reading at 25 °C | 425 nm | Adaptation of the ASTM Manual of Water and Environmental Technology, D1426 Nessler Method | HI93700-01 Reagents for 100 tests; HI93700-03 Reagents for 300 tests |
| Ammonia Low Range (13 mm Vial) | 0.00 to 3.00 mg/L (as NH-N) | 0.01 mg/L | ±0.10 mg/L or ± 5% of reading at 25 °C, whichever is greater | 425 nm | Adaptation of the ASTM Manual of Water and Environmental Technology, D1426 Nessler Method | HI93764A-25 Reagents for 25 tests |
| Ammonia Low Range ISO (13 mm Vial) | 0.000 to 2.500 mg/L (as NH+) | 0.001 mg/L | ±0.015 mg/L ± 3% of reading at 25 °C | 690 nm | ISO 23695 | HI96791-25 Reagents for 25 tests |
| Ammonia Medium Range | 0.00 to 10.00 mg/L (as NH-N) | 0.01 mg/L | ±0.05 mg/L ± 5% of reading at 25 °C, whichever is greater | 425 nm | Adaptation of the ASTM Manual of Water and Environmental Technology, D1426, Nessler Method | HI93715-01 Reagents for 100 tests; HI93715-03 Reagents for 300 tests |
| Ammonia High Range | 0.0 to 100.0 mg/L (as NH+) | 0.1 mg/L | ±0.5 mg/L ± 5% of reading at 25 °C | 425 nm | Adaptation of the ASTM Manual of Water and Environmental Technology, D1426, Nessler Method | HI93733-01 Reagents for 100 tests; HI93733-03 Reagents for 300 tests |
| Ammonia High Range (13 mm Vial) | 0.0 to 100.0 mg/L (as NH-N) | 0.1 mg/L | ±1.0 mg/L or ± 5% of reading at 25 °C, whichever is greater | 430 nm | Adaptation of the ASTM Manual of Water and Environmental Technology, D1426 Nessler Method | HI93764B-25 Reagents for 25 tests |
| Bromine | 0.00 to 10.00 mg/L (as Br) | 0.01 mg/L | ±0.08 mg/L ±3% of reading at 25 °C | 525 nm | Adaptation of the Standard Methods for the Examination of Water and Wastewater, 18th Edition, DPD Method | HI93716-01 Reagents for 100 tests; HI93716-03 Reagents for 300 tests |
| Calcium | 0 to 400 mg/L (as Ca²+) | 1 mg/L | ±10 mg/L ±5% of reading at 25 °C | 466 nm | Adaptation of the Oxalate Method | HI937521-01 Reagents for 50 tests; HI937521-03 Reagents for 150 tests |
| Calcium, Marine | 200 to 600 mg/L (as Ca²+) | 1 mg/L | ± 5% of reading at 25 °C | 610 nm | Adaptation of the Zincon Method | HI758-26 Reagents for 25 tests |
| Chloride | 0.0 to 20.0 mg/L (as Cl-) | 0.1 mg/L | ±0.5 mg/L ±5% of reading at 25 °C | 455 nm | Adaptation of the Mercury (II) Thiocyanate Method | HI93753-01 Reagents for 100 tests; HI93753-03 Reagents for 300 tests |
| Chlorine Dioxide | 0.00 to 2.00 mg/L (as ClO) | 0.01 mg/L | ±0.10 mg/L ±5% of reading at 25 °C | 575 nm | Adaptation of the Chlorophenol Red Method | HI93738-01 Reagents for 100 tests; HI93738-03 Reagents for 300 tests |
| Chlorine Dioxide (Rapid) | 0.00 to 2.00 mg/L (as ClO) | 0.01 mg/L | ±0.10 mg/L ±5% of reading at 25 °C | 525 nm | Adaptation of Standard Methods for the Examination of Water and Wastewater, 18th Edition, 4500 ClO D | HI96779-01 Reagents for 100 tests; HI96779-03 Reagents for 300 tests |
| Chlorine, Free Ultra Low Range | 0.000 to 0.500 mg/L (as Cl) | 0.001 mg/L | ±0.020 mg/L ±3% of reading at 25 °C | 525 nm | Adaptation of the EPA DPD Method 330.5 | HI95762-01 Reagents for 100 tests;HI95762-03 Reagents for 300 tests |
| Chlorine, Free Low Range (Powder Reagent) | 0.00 to 5.00 mg/L (as Cl) | 0.01 mg/L | ±0.03 mg/L ±3% of reading at 25 °C | 525 nm | Adaptation of the EPA DPD Method 330.5 | HI93701-01 Reagents for 100 tests;HI93701-03 Reagents for 300 tests |
| Chlorine, Free Low Range (Liquid Reagent) | 0.00 to 5.00 mg/L (as Cl) | 0.01 mg/L | ±0.03 mg/L ±3% of reading at 25 °C | 525 nm | Adaptation of the EPA DPD Method 330.5 | HI93701-F Reagents for 300 tests (liquid) |
| Chlorine, Free High Range | 0.00 to 10.00 mg/L (as Cl) | 0.01 mg/L | ±0.03 mg/L ±3% of reading at 25 °C | 525 nm | Adaptation of the EPA DPD Method 330.5 | HI93734-01 Reagents for 100 tests; HI93734-03 Reagents for 300 tests |
| Chlorine, Total Ultra Low Range | 0.000 to 0.500 mg/L (as Cl) | 0.001 mg/L | ±0.020 mg/L ±3% of reading at 25 °C | 525 nm | Adaptation of the EPA DPD Method 330.5 | HI95761-01 Reagents for 100 tests; HI95761-03 Reagents for 300 tests |
| Chlorine, Total Low Range (Powder Reagent) | 0.00 to 5.00 mg/L (as Cl) | 0.01 mg/L | ±0.03 mg/L ±3% of reading at 25 °C | 525 nm | Adaptation of the EPA DPD Method 330.5 | HI93711-01 Reagents for 100 tests (powder); HI93711-03 Reagents for 300 tests (powder) |
| Chlorine, Total Low Range (Liquid Reagent) | 0.00 to 5.00 mg/L (as Cl) | 0.01 mg/L | ±0.03 mg/L ±3% of reading at 25 °C | 525 nm | Adaptation of the EPA DPD Method 330.5 | HI93701-T Reagents for 300 tests (liquid) |
| Chlorine, Total High Range | 0.00 to 10.00 mg/L (as Cl) | 0.01 mg/L | ±0.03 mg/L ±3% of reading at 25 °C | 525 nm | Adaptation of the EPA DPD Method 330.5 | HI93734-01 Reagents for 100 tests; HI93734-03 Reagents for 300 tests |
| Chlorine, Total Ultra High Range | 0 to 500 mg/L (as Cl) | 1 mg/L | ±3 mg/L ±3% of reading at 25 °C | 525 nm | Adaptation of the Standard Methods for Examination of Water and Wastewater, 20th Edition, 4500-Cl | HI95771-01 Reagents for 100 tests; HI95771-03 Reagents for 300 tests |
| Chromium (VI) Low Range | 0 to 300 µg/L (as Cr(VI)) | 1 µg/L | ±10 µg/L ±4% of reading at 25 °C | 535 nm | Adaptation of the ASTM Manual of Water and Environmental Technology, D1687 Diphenylcarbohydrazide Method | HI93749-01 Reagents for 100 tests; HI93749-03 Reagents for 300 tests |
| Chromium (VI) High Range | 0 to 1000 µg/L (as Cr(VI)) | 1 µg/L | ±5 µg/L ±4% of reading at 25 °C | 535 nm | Adaptation of the ASTM Manual of Water and Environmental Technology, D1687-92, Diphenylcarbohydrazide Method | HI93723-01 Reagents for 100 tests; HI93723-03 Reagents for 300 tests |
| Chromium (VI)/Total (13 mm Vial) | 0 to 1000 µg/L (as Cr) | 1 µg/L | ±10 µg/L ± 3% of reading | 525 nm | Adaptation of the Standard Methods of the Examination of Water and Wastewater, 22nd Edition, 3500-Cr, Diphenylcarbazide Method | HI96781-25 Reagents for 25 tests |
| Chemical Oxygen Demand Low Range EPA (13 mm Vial) | 0 to 150 mg/L (as O) | 1 mg/L | ±5 mg/L or ±4% of reading at 25 °C, whichever is greater | 420 nm | Adaptation of the EPA 410.4 Approved Method for the COD Determination on Surface Waters and Wastewaters | HI93754A-25 Reagents for 25 tests |
| Chemical Oxygen Demand Low Range Mercury Free (13 mm Vial) | 0 to 150 mg/L (as O) | 1 mg/L | ±5 mg/L or ±4% of reading at 25 °C, whichever is greater | 420 nm | Dichromate Mercury Free | HI93754D-25 Reagents for 25 tests |
| Chemical Oxygen Demand Low Range ISO (13 mm Vial) | 0 to 150 mg/L (as O) | 1 mg/L | ±5 mg/L or ±4% of reading at 25 °C, whichever is greater | 420 nm | Dichromate ISO | HI93754F-25 Reagents for 25 tests |
| Chemical Oxygen Demand Medium Range EPA (13 mm Vial) | 0 to 1500 mg/L (as O) | 1 mg/L | ±15 mg/L or ±3% of reading at 25 °C, whichever is greater | 610 nm | Adaptation of the EPA 410.4 Approved Method for the COD Determination on Surface Waters and Wastewaters | HI93754B-25 Reagents for 25 tests |
| Chemical Oxygen Demand Medium Range Mercury Free (13 mm Vial) | 0 to 1500 mg/L (as O) | 1 mg/L | ±15 mg/L or ±3% of reading at 25 °C, whichever is greater | 610 nm | Dichromate Mercury Free | HI93754E-25 Reagents for 25 tests |
| Chemical Oxygen Demand Medium Range ISO (13 mm Vial) | 0 to 1000 mg/L (as O) | 1 mg/L | ±15 mg/L or ±3% of reading at 25 °C, whichever is greater | 610 nm | Dichromate ISO | HI93754G-25 Reagents for 25 tests |
| Chemical Oxygen Demand High Range EPA (13 mm Vial) | 0 to 15000 mg/L (as O) | 1 mg/L | ±150 mg/L or ±2% of reading at 25 °C, whichever is greater | 610 nm | Adaptation of the EPA 410.4 Approved Method for the COD Determination on Surface Waters and Wastewater | HI93754C-25 Reagents for 25 tests |
| Chemical Oxygen Demand Ultra High Range (13 mm Vial) | 0.0 to 60.0 ppt (as O) | 0.1 ppt | ±0.5 ppt ±3% of reading @ 25°C | 610 nm | Adaptation of the EPA 410.4 Approved Method for the COD Determination on Surface Waters and Wastewaters | HI93754J-25 Reagents for 25 tests |
| Color ADMI Low Range | 0 to 250 ADMI Pt-Co | 1 ADMI Pt-Co | ±5 ADMI Pt-Co at 25 °C | 400-700 nm | ADMI weighted ordinate Method, analogous APHA 2120F Method | |
| Color ADMI High Range | 0 to 600 ADMI Pt-Co | 1 ADMI Pt-Co | ±20 ADMI Pt-Co at 25 °C | 400-700 nm | ADMI weighted ordinate Method, analogous APHA 2120F Method | |
| Color of Water | 0 to 500 PCU (Platinum Cobalt Units) | 1 PCU | ±10 PCU ±5% of reading at 25 °C | 460 nm | Adaptation of the Standard Methods for the Examination of Water and Wastewater, 18th Edition, Colorimetric Platinum Cobalt Method | |
| Copper Low Range | 0 to 1500 µg/L (as Cu) | 1 µg/L | ±10 µg/L ±5% of reading at 25 °C | 575 nm | Adaptation of the EPA Method | HI95747-01 Reagents for 100 tests; HI95747-03 Reagents for 300 tests |
| Copper High Range | 0.00 to 5.00 mg/L (as Cu) | 0.01 mg/L | ±0.02 mg/L ±4% of reading at 25 °C | 560 nm | Adaptation of the EPA Method | HI93702-01 Reagents for 100 tests; HI93702-03 Reagents for 300 tests |
| Cyanide | 0.000 to 0.200 mg/L (as CN-) | 0.001 mg/L | ±0.005 mg/L ±3% of reading at 25 °C | 610 nm | Pyridine-Pyrazalone | HI93714-01 Reagents for 100 tests; HI93714-03 Reagents for 300 tests |
| Cyanuric Acid | 0 to100 mg/L (as CYA) | 1 mg/L | ±1 mg/L ±15% of reading at 25 °C | 525 nm | Adaptation of the Turbidimetric Method | HI93722-01 Reagents for 100 tests; HI93722-03 Reagents for 300 tests |
| Fluoride Low Range | 0.00 to 2.00 mg/L (as F-) | 0.01 mg/L | ±0.03 mg/L ±3% of reading at 25 °C | 575 nm | Adaptation of the Standard Methods for the Examination of Water and Wastewater, 18th Edition, SPADNS Method | HI93729-01 Reagents for 100 tests; HI93729-03 Reagents for 300 tests |
| Fluoride High Range | 0.0 to 20.0 mg/L (as F-) | 0.1 mg/L | ±0.5 mg/L ±3% of reading at 25 °C | 575 nm | Adaptation of the Standard Methods for the Examination of Water and Wastewater, 18th Edition, SPADNS Method | HI93739-01 Reagents for 100 tests; HI93739-03 Reagents for 300 tests |
| Hardness, Calcium | 0.00 to 2.70 mg/L (as CaCO) | 0.01 mg/L | ±0.08 mg/L ±4% of reading at 25 °C | 523 nm | Adaptation of the Standard Methods for the Examination of Water and Wastewater, 18th Edition, Calmagite Method | HI93720-01 Reagents for 100 tests; HI93720-03 Reagents for 300 tests |
| Hardness, Magnesium | 0.00 to 2.00 mg/L (as CaCO) | 0.01 mg/L | ±0.11 mg/L ±5% of reading at 25 °C | 523 nm | Adaptation of the Standard Methods for the Examination of Water and Wastewater, 18th Edition, EDTA Colorimetric Method | HI93719-01 Reagents for 100 tests; HI93719-03 Reagents for 300 tests |
| Hardness, Total Low Range | 0 to 250 mg/L (as CaCO) | 1 mg/L | ±5 mg/L ±4% of reading at 25 °C | 466 nm | Adaptation of the EPA Method 130.1 | HI93735-00 Reagents for 100 tests (LR); HI93735-0 Reagents for 300 tests (LR - 100 tests, MR - 100 tests, HR - 100 tests) |
| Hardness, Total Medium Range | 200 to 500 mg/L (as CaCO) | 1 mg/L | ±7 mg/L ±3% of reading at 25 °C | 466 nm | Adaptation of the EPA Method 130.1 | HI93735-01 Reagents for 100 tests (MR); HI93735-0 Reagents for 300 tests (LR - 100 tests, MR - 100 tests, HR - 100 tests) |
| Hardness, Total High Range | 400 to 750 mg/L (as CaCO) | 1 mg/L | ±10 mg/L ±2% of reading at 25 °C | 466 nm | Adaptation of the EPA Method 130.1 | HI93735-02 Reagents for 100 tests (HR); HI93735-0 Reagents for 300 tests (LR - 100 tests, MR - 100 tests, HR - 100 tests) |
| Hydrazine | 0 to 400 µg/L (as NH) | 1 µg/L | ±3 µg/L ±3 % of reading at 25°C | 466 nm | Adaptation of the ASTM Manual of Water and Environmental Technology, Method D1385,p- Dimethylaminobenzaldehyde Method | HI93704-01 Reagents for 100 tests; HI93704-03 Reagents for 300 tests |
| Iodine | 0.0 to 12.5 mg/L (as I) | 0.1 mg/L | ±0.1 mg/L ±5% of reading at 25 °C | 525 nm | Adaptation of the Standard Methods for the Examination of Water and Wastewater, 18th Edition, DPD Method | HI93718-01 Reagents for 100 tests; HI93718-03 Reagents for 300 tests |
| Iron Low Range | 0.000 to 1.600 mg/L (as Fe) | 0.001 mg/L | ±0.010 mg/L ±8% of reading at 25 °C | 575 nm | Adaptation of the TPTZ Method | HI93746-01 Reagents for 50 tests; HI93746-03 Reagents for 150 tests |
| Iron High Range | 0.00 to 5.00 mg/L (as Fe) | 0.01 mg/L | ±0.04 mg/L ±2% of reading at 25 °C | 525 nm | Adaptation of Standard Methods for the Examination of Water and Wastewater, 23rd Edition, 3500-Fe B, Phenanthroline Method | HI93721-01 Reagents for 100 tests; HI93721-03 Reagents for 300 tests |
| Iron (II) (Ferrous) | 0.00 to 6.00 mg/L (as Fe²+) | 0.01 mg/L | ±0.10 mg/L ±2% of reading at 25 °C | 525 nm | Adaptation of Standard Methods for the Examination of Water and Wastewater, 23rd Edition, 3500-Fe B, Phenanthroline Method | HI96776-01 Reagents for 100 tests; HI96776-03 Reagents for 300 tests |
| Iron (13 mm Vial) | 0.00 to 6.00 mg/L (as Fe) | 0.01 mg/L | ±0.10 mg/L or ±3% of reading at 25°C | 525 nm | Adaptation of Standard Methods for the Examination of Water and Wastewater, 23rd Edition, 3500-Fe B, Phenanthroline Method | HI96786-25 Reagents for 25 tests |
| Iron Total (13 mm Vial) | 0.00 to 7.00 mg/L (as Fe) | 0.01 mg/L | ±0.20 mg/L or± 3% of reading, whichever is greater | 525 nm | Adaptation of Standard Methods for the Examination of Water and Wastewater, 23rd Edition, 3500-Fe B, Phenanthroline Method | HI96778-25 Reagents for 25 tests |
| Magnesium | 0 to 150 mg/L (as Mg²+) | 1 mg/L | ±5 mg/L ±3% of reading at 25 °C | 466 nm | Adaptation of the Calmagite Method | HI937520-01 Reagents for 50 tests; HI937520-03 Reagents for 150 tests |
| Magnesium, Marine | 1000 to 1800 mg/L (as Mg²+) | 5 mg/L | ±5% of reading | 640 nm | Adaptation of Colorimetric EDTA Method using calmagite indicator | HI783-25 Reagents for 25 tests |
| Manganese Low Range | 0 to 300 µg/L (as Mn) | 1 µg/L | ±7 µg/L ±3% of reading at 25 °C | 560 nm | Adaptation of the PAN Method | HI93748-01 Reagents for 50 tests; HI93748-03 Reagents for 150 tests |
| Manganese High Range | 0.0 to 20.0 mg/L (as Mn) | 0.1 mg/L | ±0.2 mg/L ±3% of reading at 25 °C | 525 nm | Adaptation of the Standard Methods for the Examination of Water and Wastewater, 18th Edition, Periodate Method | HI93709-01 Reagents for 100 tests; HI93709-03 Reagents for 300 tests |
| Maple Syrup | 0.00 to 100.00 %T | 0.01 %T | ±3% of reading at 25 °C | 560 nm | Direct Measure | HI93703-57 Glycerol (4 pcs.), 30 mL |
| Molybdenum | 0.0 to 40.0 mg/L (as Mo+) | 0.1 mg/L | ±0.3 mg/L ±5% of reading at 25 °C | 420 nm | Adaptation of the Mercaptoacetic Acid Method | HI93730-01 Reagents for 100 tests; HI93730-03 Reagents for 300 tests |
| Nickel Low Range | 0.000 to 1.000 mg/L (as Ni) | 0.001 mg/L | ±0.010 mg/L ±7% of reading at 25 °C | 565 nm | Adaptation of the PAN Method | HI93740-01 Reagents for 50 tests; HI93740-03 Reagents for 150 tests |
| Nickel High Range | 0.00 to 7.00 ppt (as Ni) | 0.01 ppt | ±0.07 ppt±4% of reading at 25 °C | 575 nm | Adaptation of the Photometric Method | HI93726-01 Reagents for 100 tests; HI93726-03 Reagents for 300 tests |
| Nitrate | 0.0 to 30.0 mg/L (as NO¯-N) | 0.1 mg/L | ±0.5 mg/L ±10% of reading at 25 °C | 525 nm | Adaptation of the Cadmium Reduction Method | HI93728-01 Reagents for 100 tests; HI93728-03 Reagents for 300 tests |
| Nitrate Chromotropic Acid (13 mm Vial) | 0.0 to 30.0 mg/L (as N0¯-N) | 0.1 mg/L | ±1.0 mg/L or ±3% of reading at 25 °C, whichever is greater | 410 nm | Chromotropic Acid Method | HI93766-50 Reagents for 50 tests |
| Nitrate, Marine High Range | 0.0 to 75.0 mg/L (as N0¯) | 0.1 mg/L | ±2.0 mg/L ± 5 % of reading | 505 nm | Zinc Reduction Method | HI782-25 Reagents for 25 tests |
| Nitrite Low Range | 0 to 600 µg/L (as NO¯-N) | 1 µg/L | ±20 µg/L ±4% of reading at 25 °C | 480 nm | Adaptation of the EPA Diazotization Method 354.1 | HI93707-01 Reagents for 100 tests; HI93707-03 Reagents for 300 tests |
| Nitrite Low Range (13 mm Vial) | 0 to 600 µg/L (as NO¯-N) | 1 µg/L | ±10 µg/L ± 3% of reading at 25°C, whichever is greater | 525 nm | Adaptation of the Standard Method for the Examination of Water and Wastewater, 23rd Edition, 4500B Diazotization Method, Nitrogen Nitrite | HI96783-25 Reagents for 25 tests |
| Nitrite Medium Range (13 mm Vial) | 0.00 to 6.00 mg/L (as NO¯-N) | 0.01 mg/L | ±0.10 mg/L ± 3% of reading at 25°C | 525 nm | Adaptation of the Standard Method for the Examination of Water and Wastewater, 23rd Edition, 4500B Diazotization Method, Nitrogen Nitrite | HI96784-25 Reagents for 25 tests |
| Nitrite High Range | 0 to 150 mg/L (as NO¯) | 1 mg/L | ±4 mg/L ±4% of reading at 25 °C | 575 nm | Adaptation of the Ferrous Sulfate Method | HI93708-01 Reagents for 100 tests; HI93708-03 Reagents for 300 tests |
| Nitrite, Marine Ultra Low Range | 0 to 200 µg/L (as N0¯-N) | 1 µg/L | ±8 µg/L ±4% of reading at 25 °C | 480 nm | Adaptation of the EPA Diazotization Method 354.1 | HI764-25 Reagents for 25 tests |
| Nitrite, Seawater (13 mm Vial) | 0 to 600 µg/L (as N0¯-N) | 1 µg/L | ±15 µg/L ±5% of reading at 25 °C | 525 nm | Adaptation of the Standard Method for the Examination of Water and Wastewater, 23rd Edition, 4500B Diazotization Method, Nitrogen Nitrite | HI96789-25 Reagents for 25 tests |
| Nitrogen, Total Low Range (13 mm Vial) | 0.0 to 25.0 mg/L (as N) | 0.1 mg/L | ±1.0 mg/L or ±5% of reading at 25 °C, whichever is greater | 420 nm | Chromotropic Acid Method | HI93767A-50 Reagents for up to 50 tests |
| Nitrogen, Total High Range (13 mm Vial) | 10 to 150 mg/L (as N) | 1 mg/L | ±3 mg/L or ±4% of reading at 25 °C, whichever is greater | 420 nm | Chromotropic Acid Method | HI93767B-50 Reagents for up to 50 tests |
| Oxygen, Dissolved | 0.0 to 10.0 mg/L (as O) | 0.1 mg/L | ±0.4 mg/L ±3% of reading at 25 °C | 466 nm | Adaptation of the Standard Methods for the Examination of Water and Wastewater, 18th Edition, Azide Modified Winkler Method | HI93732-01 Reagents for 100 tests; HI93732-03 Reagents for 300 tests |
| Oxygen Scavengers (Carbohydrazide) | 0.00 to 1.50 mg/L (as Carbohydrazide) | 0.01 mg/L | ±0.02 mg/L ±3% of reading at 25 °C | 575 nm | Adaptation of the Iron Reduction Method | HI96773-01 Reagents for 50 tests; HI96773-03 Reagents for 150 tests |
| Oxygen Scavengers (Diethylhydroxylamine) (DEHA) | 0 to 1000 µg/L (as DEHA) | 1 µg/L | ±5 µg/L ±5% of reading at 25 °C | 575 nm | Adaptation of the Iron Reduction Method | HI96773-01 Reagents for 50 tests; HI96773-03 Reagents for 150 tests |
| Oxygen Scavengers (Hydroquinone) | 0.00 to 2.50 mg/L (as Hydroquinone) | 0.01 mg/L | ±0.04 mg/L ±3% of reading at 25 °C | 575 nm | Adaptation of Iron Reduction Method | HI96773-01 Reagents for 50 tests; HI96773-03 Reagents for 150 tests |
| Oxygen Scavengers (Isoascorbic Acid) | 0.00 to 4.50 mg/L (as Iso-Ascorbic Acid) | 0.01 mg/L | ±0.03 mg/L ±3 % of reading at 25 °C | 575 nm | Adaptation of the Iron Reduction Method | HI96773-01 Reagents for 50 tests; HI96773-03 Reagents for 150 tests |
| Ozone | 0.00 to 2.00 mg/L (as O) | 0.01 mg/L | ±0.02 mg/L ±3% of reading at 25 °C | 525 nm | Colorimetric DPD Method | HI93757-01 Reagents for 100 tests; HI93757-03 Reagents for 300 tests; HI93703-52 Reagents for 100 tests (Optional) |
| pH | 6.5 to 8.5 pH | 0.1 pH | ±0.1 pH at 25 °C | 525 nm | Adaptation of the Phenol Red Method | HI93710-01 Reagents for 100 pH tests; HI93710-03 Reagents for 300 pH tests |
| Phenols (13 mm Vial) | 0.00 to 5.00 mg/L | 0.01 mg/L | ±0.05 mg/L ±3 % of reading at 25 °C | 510 nm | Adaptation of 4aminoantipyrine Method EPA 420.1 | HI96788-25 Reagents for 25 tests |
| Phosphate Low Range | 0.00 to 2.50 mg/L (as PO³¯) | 0.01 mg/L | ±0.04 mg/L ±4% of reading at 25 °C | 610 nm | Adaptation of the Ascorbic Acid Method | HI93713-01 Reagents for 100 tests; HI93713-03 Reagents for 300 tests |
| Phosphate High Range | 0.0 to 30.0 mg/L (as PO³¯) | 0.1 mg/L | ±1.0 mg/L ±4% of reading at 25 °C | 525 nm | Adaptation of the Standard Methods for the Examination of Water and Wastewater, 18th Edition, Amino Acid Method | HI93717-01 Reagents for 100 tests; HI93717-03 Reagents for 300 tests |
| Phosphorus, Acid Hydrolyzable (13 mm Vial) | 0.00 to 1.60 mg/L (as P) | 0.01 mg/L | ±0.05 mg/L or ±5% of reading at 25 °C, whichever is greater | 610 nm | Adaptation of the EPA method 365.2 and Standard Methods for the Examination of Water and Wastewater, 20th Edition, 4500-P E, Ascorbic Acid Method | HI93758B-50 Reagents for 50 tests |
| Phosphorus, Reactive Low Range (13 mm Vial) | 0.00 to 1.60 mg/L (as P) | 0.01 mg/L | ±0.05 mg/L or ±4% of reading at 25 °C, whichever is greater | 610 nm | Adaptation of the EPA method 365.2 and Standard Methods for the Examination of Water and Wastewater, 20th Edition, 4500-P E, Ascorbic Acid Method | HI93758A-50 Reagents for 50 tests |
| Phosphorus, Reactive High Range (13 mm Vial) | 0.0 to 32.6 mg/L (as P) | 0.1 mg/L | ±0.5 mg/L or ±4% of reading at 25 °C, whichever is greater | 420 nm | Adaptation of the Standard Methods for the Examination of Water and Wastewater, 20th Edition, 4500-P C, Vanadomolybdophosphoric Acid Method | HI93763A-50 Reagents for up to 50 tests |
| Phosphorus, Total Low Range (13 mm Vial) | 0.00 to 1.60 mg/L (as P) | 0.01 mg/L | ±0.05 mg/L or ±5% of reading at 25 °C, whichever is greater | 610 nm | Adaptation of the EPA method 365.2 and Standard Methods for the Examination of Water and Wastewater, 20th Edition, 4500-P E, Ascorbic Acid Method | HI93758C-50 Reagents for 50 tests |
| Phosphorus, Total High Range (13 mm Vial) | 0.0 to 32.6 mg/L (as P) | 0.1 mg/L | ±0.5 mg/L or ±5% of reading at 25 °C, whichever is greater | 420 nm | Adaptation of the Standard Methods for the Examination of Water and Wastewater, 20th Edition, 4500-P C, Vanadomolybdophosphoric Acid Method | HI93763B-50 Reagents for up to 50 tests |
| Phosphorus, Marine Ultra Low Range | 0 to 200 µg/L (as P) | 1 µg/L | ±5 µg/L ±5% of reading at 25 °C | 610 nm | Adaptation of Standard Methods for the Examination of Water and Wastewater, 20th Edition, Ascorbic Acid Method | HI736-25 Reagents for 25 tests |
| Potassium Low Range | 0.0 to 20.0 mg/L (as K) | 0.1 mg/L | ±2 mg/L ±7% of reading at 25 °C | 466 nm | Adaptation of the Turbidimetric Tetraphenylborate Method | HI93750-01 Reagents for 100 tests; HI93750-03 Reagents for 300 tests |
| Potassium Medium Range | 10 to 100 mg/L (as K) | 1 mg/L | ±10 mg/L ±7% of reading at 25 °C | 466 nm | Adaptation of the Turbidimetric Tetraphenylborate Method | HI93750-01 Reagents for 100 tests; HI93750-03 Reagents for 300 tests |
| Potassium High Range | 20 to 200 mg/L (as K) | 1 mg/L | ±20 mg/L ±7% of reading at 25 °C | 466 nm | Adaptation of the Turbidimetric Tetraphenylborate Method | HI93750-01 Reagents for 100 tests; HI93750-03 Reagents for 300 tests |
| Silica Low Range | 0.00 to 2.00 mg/L (as SiO) | 0.01 mg/L | ±0.03 mg/L ±5% of reading at 25 °C | 610 nm | Adaptation of the ASTM Manual of Water and Environmental Technology, D859, Heteropoly Molybdenum Blue Method | HI93705-01 Reagents for 100 tests; HI93705-03 Reagents for 300 tests |
| Silica High Range | 0 to 200 mg/L (as SiO) | 1 mg/L | ±1 mg/L ±5% of reading at 25 °C | 466 nm | Adaptation of the EPA Method 370.1 for Drinking, Surface and Saline Waters, Domestic and Industrial Wastes and Standard Method 4500-SiO | HI96770-01 Reagents for 100 tests; HI96770-03 Reagents for 300 tests |
| Silver | 0.000 to 1.000 mg/L (as Ag) | 0.001 mg/L | ±0.020 mg/L ±5% of reading at 25 °C | 570 nm | Adaptation of the PAN Method | HI93737-01 Reagents for 50 tests; HI93737-03 Reagents for 150 tests |
| Sulfate | 0 to 150 mg/L (as SO²¯) | 1 mg/L | ±5 mg/L ±3% of reading at 25 °C | 466 nm | Sulfate is precipitated with barium chloride crystals | HI93751-01 Reagents for 100 tests; HI93751-03 Reagents for 300 tests |
| Surfactants, Anionic | 0.00 to 3.50 mg/L (as SDBS) | 0.01 mg/L | ±0.04 mg/L ±3% of reading at 25 °C | 610 nm | Adaptation of the EPA Method 425.1 and Standard Methods for the Examination of Water and Wastewater, 20th Edition, 5540C, Anionic Surfactants as MBAS | HI95769-01 Reagents for 40 tests |
| Surfactants, Anionic (13 mm Vial) | 0.00 to 3.50 mg/L (as SDBS) | 0.01 mg/L | ±0.10 mg/L ±5% of reading at 25 °C | 610 nm | Adaptation of the Standard Method for the Examination of Water and Wastewater, 23rd Edition, 5540C, Anionic Surfactants as MBAS | HI96782-25 Reagents for 25 tests |
| Surfactants, Cationic (13 mm Vial) | 0.00 to 2.50 mg/L (as CTAB) | 0.01 mg/L | ±0.15 ppm ±3% of reading at 25°C | 420 nm | Bromophenol Blue Method | HI96785-25 Reagents for 25 tests |
| Surfactants, Nonionic (13 mm Vial) | 0.00 to 6.00 mg/L (TRITON X-100) | 0.01 mg/L | ±0.10 mg/L ±5% of reading at 25 °C | 610 nm | TBPE Method | HI96780-25 Reagents for 25 tests |
| Zinc | 0.00 to 3.00 mg/L (as Zn) | 0.01 mg/L | ±0.03 mg/L ±3% of reading at 25 °C | 620 nm | Adaptation of the Standard Methods for the Examination of Water and Wastewater, 18th Edition, Zincon Method | HI93731-01 Reagents for 100 tests; HI93731-03 Reagents for 300 tests |
To request a quote, speak with one of our technical experts, or learn more about the HI802, please complete the form below.