Hanna Instruments New Zealand
The Hanna Advantage: A Worldwide Leader in Technology and Innovation 100% Quality 360 Value

HI802

New visible
spectrophotometer

With barcode reader, electronic zero, vial rotation, average function

Vial barcode reader

The HI802 automatically identifies barcoded sample vials, detecting method type and range to minimize errors and streamline measurements.

Vial rotation

Vial rotation during measurement enables method identification and multiple absorbance readings, which are converted to concentration units and displayed on the LCD. This rotational averaging improves method accuracy.

Shared single-zero measurement

Single-zero measurement with DI water blank correction ensures stable readings and reflects sample features, not instrument changes, for ease of use.

Why Hanna Instruments Spectrophotometer

All commercially available spectrophotometers take a single reading, except for one competitor, which takes an average reading based on 10 absorbance readings. Hanna Instruments has raised the average to 256 readings!

AVERAGE function

Know more or to get a quote

Complete the form below

By submitting this form, I agree to the Terms and Conditions and Privacy Policy available on the Hanna Instruments Australia website.

The convenience of pre-programmed methods with the option to expand.

The HI802 comes with over 100 pre-programmed methods for chemical analysis. Methods can be updated via computer or flash drive, and up to 150 can be saved. It also provides reagent codes for easy ordering.

Easily access your favorite methods directly from the home screen to save time.

Program up to 100 custom methods with up to 10 calibration points, 5 simultaneous wavelengths, and 5 reaction timers. This flexibility allows for extensive customization beyond factory presets, accommodating specific needs and modifications.

Up to 250 methods

Barcode methods

29 methods with pre-dosed vials with barcode

Automatic method recognition

Factory methods

Up to 150 factory methods

104 methods pre-programmed

Updateable via USB key

User methods

Up to 100 user methods

The tool can be customised to suit your needs

Create new

Possibility of creating a new method with the AVERAGE reading function

Favorite methods

Quick access to the most used methods

Parameters

ParameterRangeResolutionAccuracyWavelengthMethodReagent Code
Alkalinity0 to 500 mg/L (as CaCO‚ƒ)1 mg/L±5 mg/L ±5% of reading at 25 °C610 nmBromocresol GreenHI775-26 Reagents for 25 tests
Alkalinity, Marine0 to 300 mg/L (as CaCO‚ƒ)1 mg/L±5 mg/L ±5% of reading at 25 °C610 nmBromocresol GreenHI755-26 Reagents for 25 tests
Aluminum0.00 to 1.00 mg/L (as Al³+)0.01 mg/L±0.04 mg/L ±4% of reading at 25 °C530 nmAdaptation of the Aluminon MethodHI93712-01 Reagents for 100 tests; HI93712-03 Reagents for 300 tests
Ammonia Low Range0.00 to 3.00 mg/L (as NH‚ƒ-N)0.01 mg/L±0.04 mg/L ±4% of reading at 25 °C425 nmAdaptation of the ASTM Manual of Water and Environmental Technology, D1426 Nessler MethodHI93700-01 Reagents for 100 tests; HI93700-03 Reagents for 300 tests
Ammonia Low Range (13 mm Vial)0.00 to 3.00 mg/L (as NH‚ƒ-N)0.01 mg/L±0.10 mg/L or ± 5% of reading at 25 °C, whichever is greater425 nmAdaptation of the ASTM Manual of Water and Environmental Technology, D1426 Nessler MethodHI93764A-25 Reagents for 25 tests
Ammonia Low Range ISO (13 mm Vial) 0.000 to 2.500 mg/L (as NH‚„+)0.001 mg/L±0.015 mg/L ± 3% of reading at 25 °C690 nmISO 23695HI96791-25 Reagents for 25 tests
Ammonia Medium Range0.00 to 10.00 mg/L (as NH‚ƒ-N)0.01 mg/L±0.05 mg/L ± 5% of reading at 25 °C, whichever is greater425 nmAdaptation of the ASTM Manual of Water and Environmental Technology, D1426, Nessler MethodHI93715-01 Reagents for 100 tests; HI93715-03 Reagents for 300 tests
Ammonia High Range0.0 to 100.0 mg/L (as NH‚„+)0.1 mg/L±0.5 mg/L ± 5% of reading at 25 °C425 nmAdaptation of the ASTM Manual of Water and Environmental Technology, D1426, Nessler MethodHI93733-01 Reagents for 100 tests; HI93733-03 Reagents for 300 tests
Ammonia High Range (13 mm Vial)0.0 to 100.0 mg/L (as NH‚ƒ-N)0.1 mg/L±1.0 mg/L or ± 5% of reading at 25 °C, whichever is greater430 nmAdaptation of the ASTM Manual of Water and Environmental Technology, D1426 Nessler MethodHI93764B-25 Reagents for 25 tests
Bromine0.00 to 10.00 mg/L (as Br‚‚)0.01 mg/L±0.08 mg/L ±3% of reading at 25 °C525 nmAdaptation of the Standard Methods for the Examination of Water and Wastewater, 18th Edition, DPD MethodHI93716-01 Reagents for 100 tests; HI93716-03 Reagents for 300 tests
Calcium 0 to 400 mg/L (as Ca²+)1 mg/L±10 mg/L ±5% of reading at 25 °C466 nmAdaptation of the Oxalate MethodHI937521-01 Reagents for 50 tests; HI937521-03 Reagents for 150 tests
Calcium, Marine200 to 600 mg/L (as Ca²+)1 mg/L± 5% of reading at 25 °C610 nmAdaptation of the Zincon MethodHI758-26 Reagents for 25 tests
Chloride0.0 to 20.0 mg/L (as Cl-)0.1 mg/L±0.5 mg/L ±5% of reading at 25 °C455 nmAdaptation of the Mercury (II) Thiocyanate MethodHI93753-01 Reagents for 100 tests; HI93753-03 Reagents for 300 tests
Chlorine Dioxide0.00 to 2.00 mg/L (as ClO‚‚)0.01 mg/L±0.10 mg/L ±5% of reading at 25 °C575 nmAdaptation of the Chlorophenol Red MethodHI93738-01 Reagents for 100 tests; HI93738-03 Reagents for 300 tests
Chlorine Dioxide (Rapid)0.00 to 2.00 mg/L (as ClO‚‚)0.01 mg/L±0.10 mg/L ±5% of reading at 25 °C525 nmAdaptation of Standard Methods for the Examination of Water and Wastewater, 18th Edition, 4500 ClO‚‚ DHI96779-01 Reagents for 100 tests; HI96779-03 Reagents for 300 tests
Chlorine, Free Ultra Low Range0.000 to 0.500 mg/L (as Cl‚‚)0.001 mg/L±0.020 mg/L ±3% of reading at 25 °C525 nmAdaptation of the EPA DPD Method 330.5HI95762-01 Reagents for 100 tests;HI95762-03 Reagents for 300 tests
Chlorine, Free Low Range (Powder Reagent)0.00 to 5.00 mg/L (as Cl‚‚)0.01 mg/L±0.03 mg/L ±3% of reading at 25 °C525 nmAdaptation of the EPA DPD Method 330.5HI93701-01 Reagents for 100 tests;HI93701-03 Reagents for 300 tests
Chlorine, Free Low Range (Liquid Reagent)0.00 to 5.00 mg/L (as Cl‚‚)0.01 mg/L±0.03 mg/L ±3% of reading at 25 °C525 nmAdaptation of the EPA DPD Method 330.5HI93701-F Reagents for 300 tests (liquid)
Chlorine, Free High Range0.00 to 10.00 mg/L (as Cl‚‚)0.01 mg/L±0.03 mg/L ±3% of reading at 25 °C525 nmAdaptation of the EPA DPD Method 330.5HI93734-01 Reagents for 100 tests; HI93734-03 Reagents for 300 tests
Chlorine, Total Ultra Low Range0.000 to 0.500 mg/L (as Cl‚‚)0.001 mg/L±0.020 mg/L ±3% of reading at 25 °C525 nmAdaptation of the EPA DPD Method 330.5HI95761-01 Reagents for 100 tests; HI95761-03 Reagents for 300 tests
Chlorine, Total Low Range (Powder Reagent)0.00 to 5.00 mg/L (as Cl‚‚)0.01 mg/L±0.03 mg/L ±3% of reading at 25 °C525 nmAdaptation of the EPA DPD Method 330.5HI93711-01 Reagents for 100 tests (powder); HI93711-03 Reagents for 300 tests (powder)
Chlorine, Total Low Range (Liquid Reagent)0.00 to 5.00 mg/L (as Cl‚‚)0.01 mg/L±0.03 mg/L ±3% of reading at 25 °C525 nmAdaptation of the EPA DPD Method 330.5HI93701-T Reagents for 300 tests (liquid)
Chlorine, Total High Range0.00 to 10.00 mg/L (as Cl‚‚)0.01 mg/L±0.03 mg/L ±3% of reading at 25 °C525 nmAdaptation of the EPA DPD Method 330.5HI93734-01 Reagents for 100 tests; HI93734-03 Reagents for 300 tests
Chlorine, Total Ultra High Range0 to 500 mg/L (as Cl‚‚)1 mg/L±3 mg/L ±3% of reading at 25 °C525 nmAdaptation of the Standard Methods for Examination of Water and Wastewater, 20th Edition, 4500-ClHI95771-01 Reagents for 100 tests; HI95771-03 Reagents for 300 tests 
Chromium (VI) Low Range0 to 300 µg/L (as Cr(VI))1 µg/L±10 µg/L ±4% of reading at 25 °C535 nmAdaptation of the ASTM Manual of Water and Environmental Technology, D1687 Diphenylcarbohydrazide MethodHI93749-01 Reagents for 100 tests; HI93749-03 Reagents for 300 tests
Chromium (VI) High Range0 to 1000 µg/L (as Cr(VI))1 µg/L±5 µg/L ±4% of reading at 25 °C535 nmAdaptation of the ASTM Manual of Water and Environmental Technology, D1687-92, Diphenylcarbohydrazide MethodHI93723-01 Reagents for 100 tests; HI93723-03 Reagents for 300 tests
Chromium (VI)/Total (13 mm Vial)0 to 1000 µg/L (as Cr)1 µg/L±10 µg/L ± 3% of reading525 nmAdaptation of the Standard Methods of the Examination of Water and Wastewater, 22nd Edition, 3500-Cr, Diphenylcarbazide MethodHI96781-25 Reagents for 25 tests
Chemical Oxygen Demand Low Range EPA (13 mm Vial)0 to 150 mg/L (as O‚‚)1 mg/L±5 mg/L or ±4% of reading at 25 °C, whichever is greater420 nmAdaptation of the EPA 410.4 Approved Method for the COD Determination on Surface Waters and WastewatersHI93754A-25 Reagents for 25 tests
Chemical Oxygen Demand Low Range Mercury Free (13 mm Vial)0 to 150 mg/L (as O‚‚)1 mg/L±5 mg/L or ±4% of reading at 25 °C, whichever is greater420 nmDichromate Mercury FreeHI93754D-25 Reagents for 25 tests
Chemical Oxygen Demand Low Range ISO (13 mm Vial)0 to 150 mg/L (as O‚‚)1 mg/L±5 mg/L or ±4% of reading at 25 °C, whichever is greater420 nmDichromate ISOHI93754F-25 Reagents for 25 tests
Chemical Oxygen Demand Medium Range EPA (13 mm Vial)0 to 1500 mg/L (as O‚‚)1 mg/L±15 mg/L or ±3% of reading at 25 °C, whichever is greater610 nmAdaptation of the EPA 410.4 Approved Method for the COD Determination on Surface Waters and WastewatersHI93754B-25 Reagents for 25 tests
Chemical Oxygen Demand Medium Range Mercury Free (13 mm Vial)0 to 1500 mg/L (as O‚‚)1 mg/L±15 mg/L or ±3% of reading at 25 °C, whichever is greater610 nmDichromate Mercury FreeHI93754E-25 Reagents for 25 tests
Chemical Oxygen Demand Medium Range ISO (13 mm Vial)0 to 1000 mg/L (as O‚‚)1 mg/L±15 mg/L or ±3% of reading at 25 °C, whichever is greater610 nmDichromate ISOHI93754G-25 Reagents for 25 tests
Chemical Oxygen Demand High Range EPA (13 mm Vial)0 to 15000 mg/L (as O‚‚)1 mg/L±150 mg/L or ±2% of reading at 25 °C, whichever is greater610 nmAdaptation of the EPA 410.4 Approved Method for the COD Determination on Surface Waters and WastewaterHI93754C-25 Reagents for 25 tests
Chemical Oxygen Demand Ultra High Range (13 mm Vial)0.0 to 60.0 ppt (as O‚‚)0.1 ppt±0.5 ppt ±3% of reading @ 25°C610 nmAdaptation of the EPA 410.4 Approved Method for the COD Determination on Surface Waters and WastewatersHI93754J-25 Reagents for 25 tests
Color ADMI Low Range0 to 250 ADMI Pt-Co1 ADMI Pt-Co±5 ADMI Pt-Co at 25 °C400-700 nmADMI weighted ordinate Method, analogous APHA 2120F Method€”
Color ADMI High Range0 to 600 ADMI Pt-Co1 ADMI Pt-Co±20 ADMI Pt-Co at 25 °C400-700 nmADMI weighted ordinate Method, analogous APHA 2120F Method€”
Color of Water0 to 500 PCU (Platinum Cobalt Units)1 PCU±10 PCU ±5% of reading at 25 °C460 nmAdaptation of the Standard Methods for the Examination of Water and Wastewater, 18th Edition, Colorimetric Platinum Cobalt Method€”
Copper Low Range0 to 1500 µg/L (as Cu)1 µg/L±10 µg/L ±5% of reading at 25 °C575 nmAdaptation of the EPA MethodHI95747-01 Reagents for 100 tests; HI95747-03 Reagents for 300 tests
Copper High Range0.00 to 5.00 mg/L (as Cu)0.01 mg/L±0.02 mg/L ±4% of reading at 25 °C560 nmAdaptation of the EPA MethodHI93702-01 Reagents for 100 tests; HI93702-03 Reagents for 300 tests
Cyanide0.000 to 0.200 mg/L (as CN-)0.001 mg/L±0.005 mg/L ±3% of reading at 25 °C610 nmPyridine-PyrazaloneHI93714-01 Reagents for 100 tests; HI93714-03 Reagents for 300 tests
Cyanuric Acid0 to100 mg/L (as CYA)1 mg/L±1 mg/L ±15% of reading at 25 °C525 nmAdaptation of the Turbidimetric MethodHI93722-01 Reagents for 100 tests; HI93722-03 Reagents for 300 tests
Fluoride Low Range0.00 to 2.00 mg/L (as F-)0.01 mg/L±0.03 mg/L ±3% of reading at 25 °C575 nmAdaptation of the Standard Methods for the Examination of Water and Wastewater, 18th Edition, SPADNS MethodHI93729-01 Reagents for 100 tests; HI93729-03 Reagents for 300 tests
Fluoride High Range0.0 to 20.0 mg/L (as F-)0.1 mg/L±0.5 mg/L ±3% of reading at 25 °C575 nmAdaptation of the Standard Methods for the Examination of Water and Wastewater, 18th Edition, SPADNS MethodHI93739-01 Reagents for 100 tests; HI93739-03 Reagents for 300 tests 
Hardness, Calcium0.00 to 2.70 mg/L (as CaCO‚ƒ)0.01 mg/L±0.08 mg/L ±4% of reading at 25 °C523 nmAdaptation of the Standard Methods for the Examination of Water and Wastewater, 18th Edition, Calmagite MethodHI93720-01 Reagents for 100 tests; HI93720-03 Reagents for 300 tests
Hardness, Magnesium0.00 to 2.00 mg/L (as CaCO‚ƒ)0.01 mg/L±0.11 mg/L ±5% of reading at 25 °C523 nmAdaptation of the Standard Methods for the Examination of Water and Wastewater, 18th Edition, EDTA Colorimetric MethodHI93719-01 Reagents for 100 tests; HI93719-03 Reagents for 300 tests
Hardness, Total Low Range0 to 250 mg/L (as CaCO‚ƒ)1 mg/L±5 mg/L ±4% of reading at 25 °C466 nmAdaptation of the EPA Method 130.1HI93735-00 Reagents for 100 tests (LR); HI93735-0 Reagents for 300 tests (LR - 100 tests, MR - 100 tests, HR - 100 tests)
Hardness, Total Medium Range200 to 500 mg/L (as CaCO‚ƒ)1 mg/L±7 mg/L ±3% of reading at 25 °C466 nmAdaptation of the EPA Method 130.1HI93735-01 Reagents for 100 tests (MR); HI93735-0 Reagents for 300 tests (LR - 100 tests, MR - 100 tests, HR - 100 tests)
Hardness, Total High Range400 to 750 mg/L (as CaCO‚ƒ)1 mg/L±10 mg/L ±2% of reading at 25 °C466 nmAdaptation of the EPA Method 130.1HI93735-02 Reagents for 100 tests (HR); HI93735-0 Reagents for 300 tests (LR - 100 tests, MR - 100 tests, HR - 100 tests)
Hydrazine0 to 400 µg/L (as N‚‚H‚„)1 µg/L±3 µg/L ±3 % of reading at 25°C466 nmAdaptation of the ASTM Manual of Water and Environmental Technology, Method D1385,p- Dimethylaminobenzaldehyde MethodHI93704-01 Reagents for 100 tests; HI93704-03 Reagents for 300 tests
Iodine0.0 to 12.5 mg/L (as I‚‚)0.1 mg/L±0.1 mg/L ±5% of reading at 25 °C525 nmAdaptation of the Standard Methods for the Examination of Water and Wastewater, 18th Edition, DPD MethodHI93718-01 Reagents for 100 tests; HI93718-03 Reagents for 300 tests
Iron Low Range0.000 to 1.600 mg/L (as Fe)0.001 mg/L±0.010 mg/L ±8% of reading at 25 °C575 nmAdaptation of the TPTZ MethodHI93746-01 Reagents for 50 tests; HI93746-03 Reagents for 150 tests
Iron High Range0.00 to 5.00 mg/L (as Fe)0.01 mg/L±0.04 mg/L ±2% of reading at 25 °C525 nmAdaptation of Standard Methods for the Examination of Water and Wastewater, 23rd Edition, 3500-Fe B, Phenanthroline MethodHI93721-01 Reagents for 100 tests; HI93721-03 Reagents for 300 tests
Iron (II) (Ferrous)0.00 to 6.00 mg/L (as Fe²+)0.01 mg/L±0.10 mg/L ±2% of reading at 25 °C525 nmAdaptation of Standard Methods for the Examination of Water and Wastewater, 23rd Edition, 3500-Fe B, Phenanthroline MethodHI96776-01 Reagents for 100 tests; HI96776-03 Reagents for 300 tests
Iron (13 mm Vial)0.00 to 6.00 mg/L (as Fe)0.01 mg/L±0.10 mg/L or ±3% of reading at 25°C525 nmAdaptation of Standard Methods for the Examination of Water and Wastewater, 23rd Edition, 3500-Fe B, Phenanthroline MethodHI96786-25 Reagents for 25 tests
Iron Total (13 mm Vial)0.00 to 7.00 mg/L (as Fe)0.01 mg/L±0.20 mg/L or± 3% of reading, whichever is greater525 nmAdaptation of Standard Methods for the Examination of Water and Wastewater, 23rd Edition, 3500-Fe B, Phenanthroline MethodHI96778-25 Reagents for 25 tests
Magnesium0 to 150 mg/L (as Mg²+)1 mg/L±5 mg/L ±3% of reading at 25 °C466 nmAdaptation of the Calmagite MethodHI937520-01 Reagents for 50 tests; HI937520-03 Reagents for 150 tests
Magnesium, Marine1000 to 1800 mg/L (as Mg²+)5 mg/L±5% of reading640 nmAdaptation of Colorimetric EDTA Method using calmagite indicatorHI783-25 Reagents for 25 tests
Manganese Low Range0 to 300 µg/L (as Mn)1 µg/L±7 µg/L ±3% of reading at 25 °C560 nmAdaptation of the PAN MethodHI93748-01 Reagents for 50 tests; HI93748-03 Reagents for 150 tests
Manganese High Range0.0 to 20.0 mg/L (as Mn)0.1 mg/L±0.2 mg/L ±3% of reading at 25 °C525 nmAdaptation of the Standard Methods for the Examination of Water and Wastewater, 18th Edition, Periodate MethodHI93709-01 Reagents for 100 tests; HI93709-03 Reagents for 300 tests
Maple Syrup0.00 to 100.00 %T0.01 %T±3% of reading at 25 °C560 nmDirect MeasureHI93703-57 Glycerol (4 pcs.), 30 mL
Molybdenum0.0 to 40.0 mg/L (as Mo+)0.1 mg/L±0.3 mg/L ±5% of reading at 25 °C420 nmAdaptation of the Mercaptoacetic Acid MethodHI93730-01 Reagents for 100 tests; HI93730-03 Reagents for 300 tests
Nickel Low Range0.000 to 1.000 mg/L (as Ni)0.001 mg/L±0.010 mg/L ±7% of reading at 25 °C565 nmAdaptation of the PAN MethodHI93740-01 Reagents for 50 tests; HI93740-03 Reagents for 150 tests
Nickel High Range0.00 to 7.00 ppt (as Ni)0.01 ppt±0.07 ppt±4% of reading at 25 °C575 nmAdaptation of the Photometric MethodHI93726-01 Reagents for 100 tests; HI93726-03 Reagents for 300 tests
Nitrate0.0 to 30.0 mg/L (as NO‚ƒ¯-N)0.1 mg/L±0.5 mg/L ±10% of reading at 25 °C525 nmAdaptation of the Cadmium Reduction MethodHI93728-01 Reagents for 100 tests; HI93728-03 Reagents for 300 tests 
Nitrate Chromotropic Acid (13 mm Vial)0.0 to 30.0 mg/L (as N0‚ƒ¯-N)0.1 mg/L±1.0 mg/L or ±3% of reading at 25 °C, whichever is greater410 nmChromotropic Acid MethodHI93766-50 Reagents for 50 tests
Nitrate, Marine High Range0.0 to 75.0 mg/L (as N0‚ƒ¯)0.1 mg/L±2.0 mg/L ± 5 % of reading505 nmZinc Reduction MethodHI782-25 Reagents for 25 tests
Nitrite Low Range0 to 600 µg/L (as NO‚‚¯-N)1 µg/L±20 µg/L ±4% of reading at 25 °C480 nmAdaptation of the EPA Diazotization Method 354.1HI93707-01 Reagents for 100 tests; HI93707-03 Reagents for 300 tests
Nitrite Low Range (13 mm Vial)0 to 600 µg/L (as NO‚‚¯-N)1 µg/L±10 µg/L ± 3% of reading at 25°C, whichever is greater525 nmAdaptation of the Standard Method for the Examination of Water and Wastewater, 23rd Edition, 4500B Diazotization Method, Nitrogen NitriteHI96783-25 Reagents for 25 tests
Nitrite Medium Range (13 mm Vial)0.00 to 6.00 mg/L (as NO‚‚¯-N)0.01 mg/L±0.10 mg/L ± 3% of reading at 25°C525 nm Adaptation of the Standard Method for the Examination of Water and Wastewater, 23rd Edition, 4500B Diazotization Method, Nitrogen NitriteHI96784-25 Reagents for 25 tests
Nitrite High Range0 to 150 mg/L (as NO‚‚¯)1 mg/L±4 mg/L ±4% of reading at 25 °C575 nmAdaptation of the Ferrous Sulfate MethodHI93708-01 Reagents for 100 tests; HI93708-03 Reagents for 300 tests
Nitrite, Marine Ultra Low Range0 to 200 µg/L (as N0‚‚¯-N)1 µg/L±8 µg/L ±4% of reading at 25 °C480 nmAdaptation of the EPA Diazotization Method 354.1HI764-25 Reagents for 25 tests
Nitrite, Seawater (13 mm Vial)0 to 600 µg/L (as N0‚‚¯-N)1 µg/L±15 µg/L ±5% of reading at 25 °C525 nmAdaptation of the Standard Method for the Examination of Water and Wastewater, 23rd Edition, 4500B Diazotization Method, Nitrogen NitriteHI96789-25 Reagents for 25 tests
Nitrogen, Total Low Range (13 mm Vial)0.0 to 25.0 mg/L (as N)0.1 mg/L±1.0 mg/L or ±5% of reading at 25 °C, whichever is greater420 nmChromotropic Acid MethodHI93767A-50 Reagents for up to 50 tests
Nitrogen, Total High Range (13 mm Vial)10 to 150 mg/L (as N)1 mg/L±3 mg/L or ±4% of reading at 25 °C, whichever is greater420 nmChromotropic Acid MethodHI93767B-50 Reagents for up to 50 tests
Oxygen, Dissolved0.0 to 10.0 mg/L (as O‚‚)0.1 mg/L±0.4 mg/L ±3% of reading at 25 °C466 nmAdaptation of the Standard Methods for the Examination of Water and Wastewater, 18th Edition, Azide Modified Winkler MethodHI93732-01 Reagents for 100 tests; HI93732-03 Reagents for 300 tests
Oxygen Scavengers (Carbohydrazide)0.00 to 1.50 mg/L (as Carbohydrazide)0.01 mg/L±0.02 mg/L ±3% of reading at 25 °C575 nmAdaptation of the Iron Reduction MethodHI96773-01 Reagents for 50 tests; HI96773-03 Reagents for 150 tests
Oxygen Scavengers (Diethylhydroxylamine) (DEHA)0 to 1000 µg/L (as DEHA)1 µg/L±5 µg/L ±5% of reading at 25 °C575 nmAdaptation of the Iron Reduction MethodHI96773-01 Reagents for 50 tests; HI96773-03 Reagents for 150 tests
Oxygen Scavengers (Hydroquinone)0.00 to 2.50 mg/L (as Hydroquinone)0.01 mg/L±0.04 mg/L ±3% of reading at 25 °C575 nmAdaptation of Iron Reduction MethodHI96773-01 Reagents for 50 tests; HI96773-03 Reagents for 150 tests
Oxygen Scavengers (Isoascorbic Acid)0.00 to 4.50 mg/L (as Iso-Ascorbic Acid)0.01 mg/L±0.03 mg/L ±3 % of reading at 25 °C575 nmAdaptation of the Iron Reduction MethodHI96773-01 Reagents for 50 tests; HI96773-03 Reagents for 150 tests
Ozone0.00 to 2.00 mg/L (as O‚ƒ)0.01 mg/L±0.02 mg/L ±3% of reading at 25 °C525 nmColorimetric DPD MethodHI93757-01 Reagents for 100 tests; HI93757-03 Reagents for 300 tests; HI93703-52 Reagents for 100 tests (Optional)
pH6.5 to 8.5 pH0.1 pH±0.1 pH at 25 °C525 nmAdaptation of the Phenol Red MethodHI93710-01 Reagents for 100 pH tests; HI93710-03 Reagents for 300 pH tests
Phenols (13 mm Vial)0.00 to 5.00 mg/L0.01 mg/L±0.05 mg/L ±3 % of reading at 25 °C510 nmAdaptation of 4€‘aminoantipyrine Method EPA 420.1HI96788-25 Reagents for 25 tests
Phosphate Low Range0.00 to 2.50 mg/L (as PO‚„³¯)0.01 mg/L±0.04 mg/L ±4% of reading at 25 °C610 nmAdaptation of the Ascorbic Acid MethodHI93713-01 Reagents for 100 tests; HI93713-03 Reagents for 300 tests
Phosphate High Range0.0 to 30.0 mg/L (as PO‚„³¯)0.1 mg/L±1.0 mg/L ±4% of reading at 25 °C525 nmAdaptation of the Standard Methods for the Examination of Water and Wastewater, 18th Edition, Amino Acid MethodHI93717-01 Reagents for 100 tests; HI93717-03 Reagents for 300 tests
Phosphorus, Acid Hydrolyzable (13 mm Vial)0.00 to 1.60 mg/L (as P)0.01 mg/L±0.05 mg/L or ±5% of reading at 25 °C, whichever is greater610 nmAdaptation of the EPA method 365.2 and Standard Methods for the Examination of Water and Wastewater, 20th Edition, 4500-P E, Ascorbic Acid MethodHI93758B-50 Reagents for 50 tests 
Phosphorus, Reactive Low Range (13 mm Vial)0.00 to 1.60 mg/L (as P)0.01 mg/L±0.05 mg/L or ±4% of reading at 25 °C, whichever is greater610 nmAdaptation of the EPA method 365.2 and Standard Methods for the Examination of Water and Wastewater, 20th Edition, 4500-P E, Ascorbic Acid MethodHI93758A-50 Reagents for 50 tests
Phosphorus, Reactive High Range (13 mm Vial)0.0 to 32.6 mg/L (as P)0.1 mg/L±0.5 mg/L or ±4% of reading at 25 °C, whichever is greater420 nmAdaptation of the Standard Methods for the Examination of Water and Wastewater, 20th Edition, 4500-P C, Vanadomolybdophosphoric Acid MethodHI93763A-50 Reagents for up to 50 tests
Phosphorus, Total Low Range (13 mm Vial)0.00 to 1.60 mg/L (as P)0.01 mg/L±0.05 mg/L or ±5% of reading at 25 °C, whichever is greater610 nmAdaptation of the EPA method 365.2 and Standard Methods for the Examination of Water and Wastewater, 20th Edition, 4500-P E, Ascorbic Acid MethodHI93758C-50 Reagents for 50 tests
Phosphorus, Total High Range (13 mm Vial)0.0 to 32.6 mg/L (as P)0.1 mg/L±0.5 mg/L or ±5% of reading at 25 °C, whichever is greater420 nmAdaptation of the Standard Methods for the Examination of Water and Wastewater, 20th Edition, 4500-P C, Vanadomolybdophosphoric Acid MethodHI93763B-50 Reagents for up to 50 tests
Phosphorus, Marine Ultra Low Range0 to 200 µg/L (as P)1 µg/L±5 µg/L ±5% of reading at 25 °C610 nmAdaptation of Standard Methods for the Examination of Water and Wastewater, 20th Edition, Ascorbic Acid MethodHI736-25 Reagents for 25 tests
Potassium Low Range0.0 to 20.0 mg/L (as K)0.1 mg/L±2 mg/L ±7% of reading at 25 °C466 nmAdaptation of the Turbidimetric Tetraphenylborate MethodHI93750-01 Reagents for 100 tests; HI93750-03 Reagents for 300 tests
Potassium Medium Range10 to 100 mg/L (as K)1 mg/L±10 mg/L ±7% of reading at 25 °C466 nmAdaptation of the Turbidimetric Tetraphenylborate MethodHI93750-01 Reagents for 100 tests; HI93750-03 Reagents for 300 tests
Potassium High Range20 to 200 mg/L (as K)1 mg/L±20 mg/L ±7% of reading at 25 °C466 nmAdaptation of the Turbidimetric Tetraphenylborate MethodHI93750-01 Reagents for 100 tests; HI93750-03 Reagents for 300 tests
Silica Low Range0.00 to 2.00 mg/L (as SiO‚‚)0.01 mg/L±0.03 mg/L ±5% of reading at 25 °C610 nmAdaptation of the ASTM Manual of Water and Environmental Technology, D859, Heteropoly Molybdenum Blue MethodHI93705-01 Reagents for 100 tests; HI93705-03 Reagents for 300 tests
Silica High Range0 to 200 mg/L (as SiO‚‚)1 mg/L±1 mg/L ±5% of reading at 25 °C466 nmAdaptation of the EPA Method 370.1 for Drinking, Surface and Saline Waters, Domestic and Industrial Wastes and Standard Method 4500-SiO‚‚HI96770-01 Reagents for 100 tests; HI96770-03 Reagents for 300 tests
Silver0.000 to 1.000 mg/L (as Ag)0.001 mg/L±0.020 mg/L ±5% of reading at 25 °C570 nmAdaptation of the PAN MethodHI93737-01 Reagents for 50 tests; HI93737-03 Reagents for 150 tests
Sulfate0 to 150 mg/L (as SO‚„²¯)1 mg/L±5 mg/L ±3% of reading at 25 °C466 nmSulfate is precipitated with barium chloride crystalsHI93751-01 Reagents for 100 tests; HI93751-03 Reagents for 300 tests
Surfactants, Anionic0.00 to 3.50 mg/L (as SDBS)0.01 mg/L±0.04 mg/L ±3% of reading at 25 °C610 nmAdaptation of the EPA Method 425.1 and Standard Methods for the Examination of Water and Wastewater, 20th Edition, 5540C, Anionic Surfactants as MBASHI95769-01 Reagents for 40 tests
Surfactants, Anionic (13 mm Vial)0.00 to 3.50 mg/L (as SDBS)0.01 mg/L±0.10 mg/L ±5% of reading at 25 °C610 nmAdaptation of the Standard Method for the Examination of Water and Wastewater, 23rd Edition, 5540C, Anionic Surfactants as MBASHI96782-25 Reagents for 25 tests
Surfactants, Cationic (13 mm Vial)0.00 to 2.50 mg/L (as CTAB)0.01 mg/L±0.15 ppm ±3% of reading at 25°C420 nmBromophenol Blue MethodHI96785-25 Reagents for 25 tests
Surfactants, Nonionic (13 mm Vial)0.00 to 6.00 mg/L (TRITON X-100)0.01 mg/L±0.10 mg/L ±5% of reading at 25 °C610 nmTBPE MethodHI96780-25 Reagents for 25 tests
Zinc0.00 to 3.00 mg/L (as Zn)0.01 mg/L±0.03 mg/L ±3% of reading at 25 °C620 nmAdaptation of the Standard Methods for the Examination of Water and Wastewater, 18th Edition, Zincon MethodHI93731-01 Reagents for 100 tests; HI93731-03 Reagents for 300 tests

Contact Us